Flow rate: |
0 to 5 LPM |
|
O2 concentration ± 3%*: |
0-4 LPM > 90%
5 LPM 90% |
|
O2 concentration status indicator
(optional): |
Green=normal, orange=below normal, red=call technical assistance |
|
Maximum system pressure: |
1.7 bar (25 PSI) |
|
Oxygen outlet pressure: |
0.3 bar (6 PSI) |
|
Noise: |
<48dB |
|
Electrical data: |
220-240 Va.c., 50/60 Hz |
|
Energy consumption: |
Maximum 440 Watts |
|
Entrance fuse: |
220V a.c. 5x20 T 6.3A |
|
Leakage current: |
Max 0.72 mA |
|
Dimensions (HxWxD): |
450 x 370 x 580 mms. |
|
Weight: |
27 Kgs. = 59 lbs. (pounds) |
|
Optical indicators: |
ON / OFF Switch, dust filter substitution, high temperature, low
battery, therapeutic oxygen output, O2 output ready |
|
Audible alarms: |
Power failure, low therapeutic oxygen output, internal high
temperature. |
|
Pressure exchange: |
Calibration: 25 PSI |
|
Operating temperature: |
0-40 °C |
|
Filters: |
Dust filter (1), compressor inlet (1), HEPA (1) |
|
System construction: |
Modular with double insulation |
|
Movement and transport: |
Supplied on 4 wheels, lifting handle |
|
Guarantee: |
2 years or 10.000 operating hours (whichever comes first) |
|
Safety standards: |
Constructed to I.E.C. 601-1 (1988), C.E.I. 62-5 (1980), I.E.C. EN
60601-1-2 (07/2003) |
|
Device classification: |
Class II, type BF (CEI 62/5)
Class II a (Directive 93/42 CEE) |